Oxygen Generating Plant Rental: Oxygen Generating Plant With Energy Efficiency Of 0.68 To 0.83 Kwm³ To Produce Each Cubic Meter Of Oxygen With Electricity Consumption Between 8.2 ~ 10 Kw/H. Oxygen Reservoir Output Pressure 6.0 ~ 6.8 Bar, Generation Capacity Of 12 (+/- 20%) Cubic Meters Hour, With Continuous Uninterrupted Production. The System Must Have Telemetry Monitoring With Information Control Via The Internet With Information Outputs To The Hospital Institutions Internal Computers Via The Internet, Informing Operating Status And Equipment Parameters, Having Telemetry Through The Gprs System - Sms Of Production Parameters, Saturation Of The Oxygen Produced At The Time, Alarm With Production Blocking In Case Of Low Oxygen Concentration Below 92%. The Equipment Must Have Essential Accessories For Its Operation: Instantaneous Modulation Of Electrical Energy Consumption, Frequency Inverter And The Flow Of Oxygen Produced Between The Oxygen Plant And The Hospital, Having One (01) Vertical Oxygen Reservoir Of 425 Liters, ( 01) An Ultrasonic Zirconia Oxygen Analyzer, Being Integrated Into The Oxygen Generating Plant. Compressed Air Generation System For An Oxygen Generating Plant Must Be 100% Oil-Free, Free From Purges And Any Type Of Condensate, Free From Electronic And/Or Automatic Drains, Free From A Refrigeration Dryer, Free From Coal And Coal Filters Activated On The Oxygen Generation Line, Equipment Must Be As Compact As Possible, Occupying A Maximum Space Of 10 ~ 14 M² (Ten To Fourteen Square Meters) And Easy To Transport, Avoiding Prolonged Installations With Several Connections And Pipes, As Compact As Possible. Coupled To The Oxygen Plant Is An Oxygen Compressor “Cylinder Filler” With A Flow Rate Of: 2 M³/H, With An Electrical Energy Consumption Of: 2.2 Kwh, Voltage 220V Three-Phase, 60Hz, Working Pressure Of: 150 Bar. Etc..., Rental Of Medical Compressed Air Central: Generation Of Medical Compressed Air With A Capacity Of 22 M³/H, Independent System With Two High Energy Efficiency Compressors, Compatible With Nbr 12.188, Rdc So/Anvisa. With The Following Limits; 02 (%Vlv) :20.4 A21.4, Co2 (Ppm): 500, So2 (Ppm): 1, Oils (Mg/N4): 0.1, Nox (Ppm) : 2, Co (Ppm): 5, H2o (ºc): -45.5, Particles 0, Power: 110/220 Vac Control Panel: Metal, On-Off Switch, Leds, Standby Position And Gauges. (1) Set Of Miscellaneous Filters, (1) One Condensate Separator Filter, (1) One High Efficiency Coalescing Pre-Filter For General Use. Removal Of Particles Up To I P, Including Condensed Water And Oil. Maximum Oil Residual 0.5 Mg/M3 At 21 0 C. (1) One Post - Oil Removal Coalescing Filter. Removal Of Particles Up To 0.01 Μm, Including Water And Oil Aerosols. Maximum Oil Residual 0.01 Mg/M3 At 2L ºc. (1) An Activated Carbon Filter. For Removal Of Oil Vapors, Providing Maximum Remaining Oil Content ≪ 0.003 Mg/M3 (≪ 0.003Ppm) At 21 ºc. (1) An Adsorption Dryer That Employs The Properties Of Molecular Sieves Retains Molecules By Strong Physical Attraction. At This Stage, Air Is Free And In The Percentages Indicated By The Standard, Water H2o, Carbon Dioxide Co2, Sulfur Dioxide So2, And Monoxide And Nitrogen Dioxide No/I`{O2. Must Contain Two Separation Towers, With Intelligent Control Of Valves And Pressure Gauge
|