Automatic Biochemistry Analyzer Minimum Features: Pipetting Speed Of At Least 200 Photometric Tests Per Hour; - 50 Or More Simultaneous Positions With 24-Hour Cooling For The Reagents, With A Separate On And Off Switch, Allowing The Equipment To Be Turned Off And Keeping The Reagents Cooling Active; - Sample Tray With 70 Positions For Primary Tubes Or Cuvettes; - Automatic System That Washes And Drys The Cuvettes Used; - Have A Sensor On The Sample/Reagent Pipettor Needle, Which Indicates With An Audible And On-Screen Alert, In The Event Of A Lack Of Samples Or A Lack Of Reagents; - Accept Urgent Samples So That Even When There Is A Routine In Progress; - Allows Interface With The Laboratorys Operating Software; - Uses Internal/External Barcode Reader For Samples; - Have Wave Filters With At Least 9 Different Wavelengths, Being Mandatory To Cover The Range Between 340 And 800 Nm; Automatic Washing System With Detergent; The Equipment Must Have Current Registration With The National Health Surveillance Agency – Anvisa. 12-Month Warranty From The Winning Bidder, Proof Of Technical Assistance In The Capital Recife-Pe, Or Within A 150Km Radius Of The City Of Escada-Pe. Minimum 1 Year Warranty., Reverse Osmosis System The Device Must Perform 300 Tests Per Hour, Have An Open System, Refrigerated Tray, Aspiration Volume: 20 To 300 ?L (1 ?L Interval), Container: 32 Ml And 16 Ml Up To 80 Reagent Positions, Optional Barcode Reader, Up To 40 Positions Of Patients, Aspiration Volume: 2 To 30 ?L (0. 1 Μl Range), 44 Reaction Cells, Reaction Volume: 150 To 330 ?L, Fixed Cycle: 18 Seconds, Total Reaction Time: 10 Minutes, End Point And Kinetics. Biochemistry And Turbidimetry Tests, Linear And Non-Linear Calibration Photometry With Diffraction Grating With 8 Different Wavelengths (Optional With 12 Different Wavelengths), Bichromatic Reading • Halogen Lamp, Automatic Washing Of Reaction Cells, Uses Only One Solution To Wash The Cells, Water Consumption: 8 L/H, Independent Mixer, Probe For Addition Of Sample And Reagent, Level Detector For Samples And Reagents, Optional Barcode Reader. The Electrical Supply Must Be Automatic Bivolt Or In Accordance With The National Electrical Network (110 / 220V, 50/60 Hz). The Equipment Must Come With Ups For Proper Use. Minimum Characteristics: Nominal Peak Power 2000 Va / 1900 W; - Nominal Input Voltage: 120 / 220 V; - Nominal Output Voltage: 220V (Can Be Adjusted Manually); - Contain 4 Output Sockets, 3X 10A And 1X 20A; - Battery Management That Notifies You When The Battery Needs To Be Replaced; - Self Test To Verify The Initial Conditions Of The Equipment; - Visual Signaling With All Conditions Of The Equipment, Battery And Electrical Network; - Internal Leak-Proof Sealed Vrla Type Batteries. The Equipment Must Have A 12-Month Warranty By The Winning Bidder, Digital Biological Centrifuge. The Equipment Must Have An Induction Motor, With Silent, Stable And Low-Maintenance Operation, A Fixed-Angle Rotor With Adjustable Speed Of Up To 4,000 Rpm, Electronic Rotation Control, And A Minimum Capacity For 12 (Twelve) Standard Tubes Of 12 X 75 Mm, Allowing The Use Of Adapters For Other Compatible Volumes. The Equipment Must Have A Digital Panel For Adjusting And Displaying The
|