Thermodisinfector For Cleaning And Thermal Disinfection Of Instruments. (It Is Intended For The Automated Processing Of Materials Used In Healthcare Establishments, Ensuring Standardized Cleaning, Occupational Risk Reduction And Valid Thermal Disinfection In Accordance With Current Technical Standards. The Thermodisinfector Must Have An Internal Chamber Made Of Aisi 316L Stainless Steel, With A Sanitary Finish Resistant To Corrosion, And A Minimum Useful Volume Of 200 Liters. The Door Must Have A Tempered Glass Display, Sealed With A Non-Toxic Silicone Seal And Automatic Or Semi-Automatic Closing. The Equipment Must Have Automatic Cleaning, Rinsing, Thermal Disinfection And Drying Programs, Reaching A Minimum Temperature Of 90 °C ± 2 °C, Maintained For At Least 5 Minutes, With Registration And Validation Of All Cycle Parameters, Including Time, Temperature, Pressure And Faults. The Control Must Be Carried Out Using A Microprocessor System With A Graphic Screen For Viewing The Steps, Alarms And Faults, Being Capable Of Storing At Least 100 Cycles In The Internal Memory. The Equipment Must Allow The Export Of Reports Via A Usb Port Or Ethernet Connection, As Well As Having. Integrated Thermal Printer Or Interface For External Printer, Guaranteeing Process Traceability. At A Minimum, The Following Accessories Must Be Provided: 01 Universal Rack For Trays, 01 Specific Rack For Surgical Instruments, 01 Rack For Cannulated Materials With Individualized Irrigation System, Support For Glassware And Small Accessories, In Addition To Interchangeable Stainless Steel Trays. The Electrical Supply Must Have A Voltage Of 220 V Or 380 V, 60 Hz, Allowing ±10%. The Equipment Must Have Inlets For Cold And Hot Water, With A Built-In Filtering System, In Addition To An Automatic Detergent And Neutralizer Dosing System. The Equipment Must Fully Comply With Rdc 15/2012. (Anvisa), Rdc 611/2022 (Good Manufacturing Practices) And Standard A, Surgical Grade Paper Sealer For Sealing Sterilization Packaging. Ensuring Sealing Integrity Before And After Sterilization Cycles. The Equipment Must Have A Carbon Steel Structure With Anti-Corrosive Electrostatic Painting Or Stainless Steel, Resistant To Oxidation And Easy To Clean, Shielded Electrical Resistance For Easy Replacement And Protection Against Overheating – 0º To 300º C; Synchronized; Sealing Speed Of Approximately 10 M/Min; Independent Heating Control And Motor Drive. Must Have Continuous Sealing System (Roller) Or Thermal Impulse, With Minimum Sealing Strip Of 12 Mm, Minimum Sealing Width Of 250 Mm And Minimum Weld Thickness Of 1.5 Mm, Continuous And Without Flaws. It Must Have An Integrated Cutting Or Tearing Device For The Packaging After Sealing, Adjustable Temperature Control With A Visual Operating Indicator And A Maximum Heating Time Of 5 Minutes. The Equipment Must Also Have Non-Slip Feet Or Rubberized Support Casters, Ensuring Stability On The Bench. (Minimum 5 Cm And Maximum 30 Cm) And Suitable For Use With Packaging Compatible With Proce
|