Cardioverter 01 Which Must Be Composed Of A Configurable Biphasic Defibrillator/Configurable Monitor Set Intended For Defibrillation And Monitoring Vital Signs Of Adult, Pediatric And Neonatal Patients. Consisting Of A Single Body, It Has An 8.4-Inch Lcd Screen With A Very High Minimum Resolution Of 800 X 600 Pixels And An Intuitive Touch Screen Navigation System. Parameters. Maximum Weight Of 8 Kg Including The Paddles, Rechargeable Lithium-Ion Battery With A Minimum Autonomy Of 3 Hours, Must Have A Remaining Charge Indicator On The Screen. Ecg. Must Have The Ability To Store Events With The Possibility Of Reviewing The Curve Stored On The Screen (Stored Events Are Cardioversion, Defibrillation, Arrhythmia Detection And Manually Entered Events Such As Endotracheal Access, Intravenous Access, Among Others). Advanced Arrhythmias. Must Have Signaling Devices With Audible And Visual Indication, Command For Temporary Interruption Of Audible Alarms For A Few Minutes, Continuous System For Visual Alarms. (Fc), Pulse Oximetry And Non-Invasive Blood Pressure. The Biphasic Defibrillator Must Have Manual Energy Intensity Of 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50, 70, 100, 130, 150, 170, 200 And 250 J. Indicator With Text Commands, Audible Alerts, Qrs Beep, Battery Charge Level, Equipment, Fixed-Digital X-Ray Device 02Fixed, Digital Features: The General-Purpose X-Ray Device For Carrying Out Examinations Of The Pulmonary Skeleton System, Floor/Ceiling Type, Floor/Ground Or Floor/Table. Stretcher, Lying On The Back, Side, Ventral On The Table, Examinations Of The Chest, Spine, Skull And Extremities In General. Fully Digital System, Without The Need For A Chassis Or Mobile Digitizers. Microprocessor, With Variable Voltage Of At Least 40 Kv For Radiography, And Of At Least 600 Ma Or Greater, Having A Minimum Exposure Time In Radiography Of 1 Ms Up To 4S Or Greater, Automatic Exposure Control That Allows A Technique Of At Least 3 Points, As Well As Free For Bucky Table And Bucky Wall With 3 Voltage Fields (Kv), Current (Ma) And Time (S), Independent Or Together (Mas). Minimum Generator Power Of 50 Kw Or Greater, Variable Range Equal To 0.5 Mas Or Less Up To 800 Mas Or Less. Coarse Focus Of 1.2 Mm, With Minimum Focal Power Of 20 Kw Up To 50 Kw Or Greater, Tube Support, (Stand) Ti Mount
|