Computed Radrography Digital Imaging System For Printing: General Description Computed Radiography System (Cr) Monocassette Type, Intended For Digitizing Images For General Radiology, With Capacity For Reading Cassettes In At Least Three Sizes: 18 By: Digitizer Unit (Scanner Cr), Compatible Cassettes, Acquisition Station With Dedicated Software For Patient Registration And Digital Image Processing. Minimum Technical Characteristics Digitizer (Cr): Automatic Recognition Of The Size And Type Of Cassette Inserted; Minimum Processing Capacity Of 60 Cassettes/Hour In Size 35 X 43 Cm; Minimum Gray Scale Of 12 Bits, Ensuring High Image Quality And Fidelity; Monocassette Operation (Single Cassette Input).Acquisition Station (Processing Console)Minimum Hardware Required: Intel Core I3 Processor (Or Higher, Or Equivalent In Performance); 4 Gb Of Ram Memory (Or Higher); 500 Gb Hard Disk Storage (Or Higher); Touch Sensitive Monitor, Minimum 21 Inches. Image Acquisition And Processing Software Must Contain, At A Minimum, The Following Tools And Functions: Full Screen Image Viewing; Independent Adjustment Of Latitude, Contrast And Brightness; Cropping The Image According To The Size And Position Defined By The User; Insertion Of Notes: Text..., Hematology Analyzer: The Equipment Has An Approximate Weight Of 16 To 17 Kg, An Electrical Supply Of 100 To 240 V (5060 Hz) And Dimensions Of 43 36 36 Cm. It Has A Built-In Thermal Printer And An 8.4-Inch Touch-Sensitive Color Lcd Screen. It Presents Performance Of 50 To 60 Tests Per Hour, Using A Sample Volume Of Just 10 Μl In Cbc 3Diff Mode, With Stability Of Up To 48 Hours After Collection When Refrigerated. Allows Operation With Open Or Closed Tube And Has Memory For Up To 1000 Results With Histograms. The Data Management System Includes A Complete Menu With Calibration, Reagents, Maintenance And Quality Control Records, As Well As Usb And Lan Connectivity For Data Transmission And Import. The Equipment Performs 19 Hematological Parameters, Including Wbc, Rbc, Hgb, Hct, Plt And Derived Indices, With Display Of Plt, Rbc And Wbc Curves. It Presents High Precision, With A Coefficient Of Variation Of Less Than 2.5 For Wbc, 2.0 For Rbc And Hct, 1.5 For Hgb And 5.0 For Plt, Within The Established Linearity Ranges, Guaranteeing Reliable And High-Quality Results., Biochemical Analysis Equipment: The Biochemistry Equipment Performs Colorimetric, Enzymatic, Kinetic And Turbidimetric (Standard, Factor Or Standard Curve), Operating In An Open System With Up To 90 Programming Possibilities For Different Reagents. It Has A Monochromator With Seven Filters (340, 405, 505, 546, 578, 620 And 670 Nm), A Reading Range Of -0, 100 To 3, 500 Abs And A Passband Of 10 Nm To 340 Nm And 6 Nm For The Other Filters. It Presents Stray Light Less Than 0.01T And Allows Monochromatic And Bichromatic Readings. It Has A Continuous Flow Metal Cuvette (32 Μl), Thermostated Between 25 C And 37 0.1 C, With Contamination Lower Than 1. It Performs Readings In Absorbance, End Point, Kinetics, Fixed Time And Differential, With Calculations By Pattern, Multi Pattern Or Factor. Includes Quality Control With Levey-Jennings Chart, Built-In Graphic Printer And
|